Recall Z-0726-2022 ยท DSQ Ventricular (Assist) Bypass

Heartware, Inc.

HeartWare Ventricular Assist Device (HVAD) System

DateJan 27, 2022
StatusOpen, Classified
ReasonUpdated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power source connector pins, as this could remove the lubricant that is applied to the pins as a mitigation for power switching.
Root causeLabeling design
ActionStarting 27-Jan-2022 and ending 02-FEB-2022, Medtronic initiated an Urgent Medical Device Notice to global customers to inform them of the intent to update the cleaning instructions for power source connector pins on the HVAD" System Monitor AC Adapter, DC Adapter, and Batteries. The communication provided customers the updated instructions, the complaints received as well as the potential hazards of cleaning the power source connector pins. In addition to the Urgent Medical Device Notice a patient template was provided to facilitate the VAD Coordinators in notifying their patients. In the Un
Quantity197,889 units
DistributionDistribution throughout the United States. OUS distribution in Argentina, Armenia, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Kazakhstan, Korea (So
FDA recordOpen on fda.gov