Recall Z-0723-2022 ยท OBO Tomography, Optical Coherence

Carl Zeiss Meditec, Inc.

CIRRUS HD-OCT, Models 5000 AngioPlex and 6000 AngioPlex

DateJan 14, 2022
StatusOpen, Classified
ReasonOptic nerve head angiography scan to be turned off due to its distribution without pre-market clearance.
Root causeNo Marketing Application
ActionOn 01/14/22, recall notices were mailed to customers who were asked to do the following: 1. Until further notice, please discontinue the acquisition of the ONH Angiography 4.5x4.5mm Scan. 2. Using the provided software kit(s), install software version 11.5.3 on all affected instruments. You will be receiving one software kit per affected instrument according to our records. An excerpt of the IFU has been included Section 4.7 Update Instrument Software, these instructions apply to all models. Refer to the IFU Section 4 for the full installation instructions. 3. Complete and return the Acknowle
Quantity1,272 devices
DistributionU.S Nationwide Distribution.: OH, CA, IN, AR, IA, LA, WI, OK, IL, OR, AL, KY, WY, WA, MN, NE, PA, SD, AZ, PR, TN, GA, FL, DE, NY, NH, TX, MS, MA, NV, UT, WV, ME, CO, AK, NM, KS, MO, MI, MD, ID, CT, NJ, MT, NC, VA, HI, VT, ND, SC, DC, RI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181534CIRRUS HD-OCT
Carl Zeiss Meditec, Inc., Feb 15, 2019