Recall Z-0680-2022 ยท QKP Coronavirus Antigen Detection Test System.

CELLTRION USA INC

Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02

DateMar 2, 2022 (initiated Dec 2, 2021)
StatusTerminated
ReasonReturned test kits were erroneously shipped to customers. The kits are intended for Research Use Only (RUO). Additional RUO test kits were shipped to customers unlikely to use them for RUO purposes, and the firm provided a letter assuring distributors that RUO tests could be used for clinical diagnosis.
Root causeEmployee error
ActionOn December 2, 2021, the firm notified the affected distributor that the two lots are Research Use Only and should not be used. The distributor then notified the customer. Customer reported all product had been used.
Quantity8080 kits originally distributed; 726 kits redistributed
DistributionDistributed to TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.