Recall Z-0701-2022 ยท OHQ Multi-Analyte Controls Unassayed

Microbiologics Inc

Helix Elite" Molecular Standards: Inactivated Standard. Inactivated Influenza A/B and Respiratory Syncytial Virus

DateJun 6, 2017
StatusTerminated
ReasonReal-time shelf life testing failed at 24 months
Root causeProcess design
ActionMicrobiologics notified customers on about 06/06/2017 via email containing a letter. The letters indicated the affected products and instructed customers to update their records with the corrected expiration date, which was provided for each affected lot.
Quantity1 unit
DistributionWorldwide distribution - US Nationwide distribution in the states of AK, CA, DC, FL, IA, ID, KS, LA, MD, MN, MO, NE, NH, NJ, NY, OH, OK, PA, SC, TX, VA, WA, WV and the countries of Australia, Belgium, Canada, India, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.