Recall Z-0699-2022 ยท OHQ Multi-Analyte Controls Unassayed

Microbiologics Inc

Helix Elite" Molecular Standards: Inactivated Standard. Inactivated Respiratory Syncytial Virus A Strain Long. Target pathogens are lyophilized in a PCR compatible matrix, and stable at 2C-25C through the expiration date. Each lyophilized pellet is packaged in a single-use foil pouch. Each kit contains 5 unassayed lyophilized pellets, packaged in individual vials for ease-of-use and to avoid cr

DateJun 6, 2017
StatusTerminated
ReasonReal-time shelf life testing failed at 24 months
Root causeProcess design
ActionMicrobiologics notified customers on about 06/06/2017 via email containing a letter. The letters indicated the affected products and instructed customers to update their records with the corrected expiration date, which was provided for each affected lot.
Quantity1 unit
DistributionWorldwide distribution - US Nationwide distribution in the states of AK, CA, DC, FL, IA, ID, KS, LA, MD, MN, MO, NE, NH, NJ, NY, OH, OK, PA, SC, TX, VA, WA, WV and the countries of Australia, Belgium, Canada, India, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.