ClearanceBrief
Recall Z-0698-2022 ยท OHQ Multi-Analyte Controls Unassayed

Microbiologics Inc

Microbiologics QC Sets and Panels. BD MAX" CT/GC/TV Verification and Control Panel. Containing High, Medium and Low Level Controls.

DateJun 6, 2017
StatusTerminated
ReasonReal-time shelf life testing failed at 24 months
Root causeProcess design
ActionMicrobiologics notified customers on about 06/06/2017 via email containing a letter. The letters indicated the affected products and instructed customers to update their records with the corrected expiration date, which was provided for each affected lot.
Quantity29 units
DistributionWorldwide distribution - US Nationwide distribution in the states of AK, CA, DC, FL, IA, ID, KS, LA, MD, MN, MO, NE, NH, NJ, NY, OH, OK, PA, SC, TX, VA, WA, WV and the countries of Australia, Belgium, Canada, India, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.