Recall Z-0748-2022 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Siemens Medical Solutions USA, Inc.

Diagnostic Ultrasound System, ACUSON Juniper, UDI: (01)04056869152059, Model: 11335791, Rx, CE Ultrasound imaging system intended for the following applications: Vascular, Abdominal, Neonatal, Echocardiographical, Pediatric, Small Parts, Musculoskeletal, Obstetrical, Gyneological, Intraoperative, and Urological applications using different ultrasound transducers for different applications.

DateJan 19, 2022
StatusOpen, Classified
ReasonThe clip store function in the ultrasound imaging system does not work when the system has a disk full error. This could cause a delay in treatment if the ultrasound system is unable to save clips as study documentation during a high risk procedure, such as a stress echo exam.
Root causeSoftware design
ActionThe firm, Siemens Healthineers, sent "URGENT MEDICAL DEVICE SAFETY CORRECTION" Letters Juniper 1.0 (VA10D/VA10E/VA10F) to affected customers via Siemens update program US026/21/S informing them of the issue and advising them the firm will correct this issue with a free-of-charge software update to their ACUSON Juniper ultrasound system. The customers were instructed to do the following: Check the available storage on the ultrasound system prior to starting a study where clips are vital to the documentation, such as stress echo. If available, use a different ultrasound system capable of perfo
Quantity2846
DistributionWorldwide distribution: US (nationwide) and countries of: Algeria, Andorra, Argentina, Australia, Austria, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Faroe Islands, France, Germany,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201130ACUSON Juniper Diagnostic Ultrasound System
Siemens Medical Solutions USA, Inc., May 26, 2020