Recall Z-0650-2022 ยท JDR Staple, Fixation, Bone

Merete Medical GmbH

Merete PediatrOS RigidTack, REF CP20120, Size 20 mm, Sterile, Rx Only

DateJan 18, 2022
StatusTerminated
ReasonIncorrect marketing label (25 mm) was applied to product size 20 mm.
Root causeError in labeling
ActionCustomers were notified of the recall via phone call on 01/19/2022 and instructed to quarantine any devices from the recalled lot. The recalling firm also informed customers that a written notice was going to be distributed with further instructions. Devices remaining in stock should be separated and quarantined, while those implanted should have patients informed and x-rays conducted to evaluate if the appropriate size was utilized. Additionally, customers were notified via email on 01/27/2022 of the recall in an Urgent Medical Device Recall Notice. The notice identified Lot MS2117816 of Pe
Quantity26 units
DistributionDomestic: Illinois; Foreign: Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151762PediatrOS RigidTack/FlexTack
Merete Medical GmbH, Dec 28, 2015