ClearanceBrief
Recall Z-0708-2022 ยท LZU Plug, Punctum

Lacrimedics Inc

OPAQUE Herrick Lacrimal Plugs, HP3, 0.3 mm, HP5, 0.5mm, HP7, 0.7.mm, STERILE, EO, Rx only, CE Dry eye treatment

DateJan 17, 2022
StatusOpen, Classified
ReasonThe sterile pouch seal may contain channels that could affect the seal integrity and potentially compromise sterility.
Root causePackaging process control
ActionUrgent : Medical Device Recall, Collagen Plugs and Opaque Herrick Lacrimal Plugs customer letters, dated January 17, 2022 and Urgent : Medical Device Recall, Collagen Plugs and Opaque Herrick Lacrimal Plugs distributor letters, dated January 17, 2022 sent starting 1/24/2022 (through 1/28/2022 US and 2/4/2022 (OUS)) via USPS certified letter with return receipt; Distributors and customers are instructed in letter to discontinue distribution and/or use and return remaining product for replacement " Contact their customers regarding the product(s) and request the return of remaining product fo
Quantity5767 units
DistributionNationwide and Worldwide Distribution to following countries - AR. BE, BO, BR, CH, CN, CO, CZ, DE, EC, ES, FI, GE, IN, IL, MY, MX, NZ, PE, RO, SG, SA, TR, TW, TZ, UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021831OPAQUE HERRICK LACRIMAL PLUG, 0.3MM,0.5MM,0.7MM, MODELS HLP3-OPB,HLP50OPB,HLP7-OPB
Lacrimedics, Inc., Jul 1, 2002