ClearanceBrief
Recall Z-0802-2022 ยท FRO Dressing, Wound, Drug

CellEra LLC

20522 Vitale Silver calcium Alginate Dressings 2x2, 10/box

DateMar 16, 2022 (initiated Dec 15, 2021)
StatusTerminated
ReasonFirm discovered that an incorrect version of the Instructions for Use (IFU) containing language beyond the cleared 510(k) were inserted in the affected products. Firm is replacing these with the correct version of the IFU.
Root causeLabeling Change Control
ActionOn December 15, 2021, the firm notified customers who had received product from CellEra within the past six months. The firm requested these customers hold any merchandise on hand pending receipt of replacement IFU's. The firm communicated with those companies to determine how many new IFU's they require for their inventory and for their customers.
Quantity9200 ea
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.