Recall Z-0714-2022 ยท JTR Kit, Quality Control For Culture Media

Microbiologics Inc

KWIK-STIK, Quality control kit for culture media, sold as: KWIK-STIK (single-pack) Enterococcus faecalis derived from ATCC 51299

DateOct 6, 2014
StatusTerminated
ReasonThis strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
Root causeNonconforming Material/Component
ActionMicrobiologics notified customers on about 10/06/2014 via letter titled "E.faecalis Gentamicin Susceptibility Notification." Instructions included to inspect inventory for affected products. If customers have received the product identified within the recall, it is recommended that the intended use of the organism be reviewed to determine if Gentamicin synergy results have an impact on use. It is recommended to discard these lots if an impact is to be expected. Questions or concerns can be directed to the Technical Support team at 320.229.7064 or techsupport@microbiologics.com.
Quantity150 units
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia/Hrvatska, Ecuador, Finland, France, Germany, Guam, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Kenya, Republic of Korea, Latvia, Macedonia, Malaysia, Nether