ClearanceBrief
Recall Z-0627-2022 ยท QEW Peripheral Mechanical Thrombectomy With Aspiration

TELEFLEX MEDICAL INC

Arrow-Trerotola PTD Set (Percutaneous Thrombolytic Device), PT-45509 containing components: PT-45509-S6, PT-45509-T The Arrow-Trerotola Percutaneous Thrombolytic Device (PTD) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venous (AV) fistulae and synthetic dialysis grafts.

DateFeb 15, 2022 (initiated Jan 11, 2022)
StatusTerminated
ReasonThe firm received reports indicating PTD tip separation during use.
Root causeProcess change control
ActionArrow International LLC/Teleflex notified their consignees by letter on 01/11/2022. The letter explained the issue and requested that the consignee cease use and distribution. Distributors were directed to notify their customers. The firm is seeking return of the affected product. The following instructions were provided: Action list number 1 Medical facilities 1. We request that you check your inventory for product within the scope of this FSCA. Users should cease use and distribution of impacted product and quarantine immediately. 2. If you do not have stock in scope of this FSCA m
Quantity3668 units
DistributionWorldwide distribution: US Nationwide and the countries of AU, BE, CA, CL, CZ, DE, ES, FR, GB, IL, IT, KR, KW, NL, PA, PH, PT, SA, TR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011056ARROW-TREROTOLA PERCUTANEOUS THROMBOLYTIC DEVICE OR PTD
Arrow Intl., Inc., Oct 24, 2001