ClearanceBrief
Recall Z-0649-2022 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Veradius Unity, Model #718132

DateDec 2, 2021
StatusOpen, Classified
ReasonFirm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.
Root causeUnder Investigation by firm
ActionOn December 2, 2021, the firm issued Urgent Medical Device Correction letters to affected customers. The Medical Device Correction includes the following instructions for the Customer: o Ensure that the wired foot switch is always connected to the system. o Immediately start using the wired foot switch or hand switch in the event the wireless foot switch loses connection. o Perform a daily check for damage and correct functioning of the hand switch and wired foot switch as described by the User Routine Checks section in your Instruction for Use. o Circulate this notice to all users o
Quantity12 US; 44 ROW
DistributionDomestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142708VERADIUS UNITY
Philips Medical Systems Nethrlands BV, Oct 22, 2014