Recall Z-0640-2022 ยท FSA Lift, Patient, Non-Ac-Powered

Hill-Rom, Inc.

Liko Mobile Lifts: Primarily intended for use in nursing homes LikoM230 Model: 2050015

DateDec 23, 2021
StatusOpen, Classified
ReasonMissing bushing on Liko M220 and M230 causing wear and potential for patient falling
Root causeProcess control
ActionHillRom issued Urgent Medical Device Correction Letter (FSCA Identifier: FA 2021-010-002-LUL-001) via third party on 23 December 2021 via UPS Ground Service to Distributors and Direct Users. Letter states reason for recall, health risk and action to take: 1. Review the product label on your Liko M220/M230 to determine if it is impacted by this field action by identifying the manufacturing date of the device (i.e. Manufacturing date between 2016-DEC-27 and 2021 SEP-01). See figure 2. 2. Please periodically inspect your Liko device for damage or wear. Report visible wearing or damage to your Te
Quantity1215 units
DistributionWorldwide distribution - US Nationwide and the countries of Austria, Austria, Bahrain, Canada, Switzerland, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Norway, Saudi Arabia, Sweden, Slovenia, Taiwan (Province of China), South Africa, Oman.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.