ClearanceBrief
Recall Z-0589-2022 ยท JTR Kit, Quality Control For Culture Media

Microbiologics Inc

QC Sets and Panels, KWIK-STIK Blood Culture Identification (BCID) Verification Panel (Live Culture). KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms

DateFeb 8, 2022 (initiated Feb 18, 2020)
StatusTerminated
ReasonTests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
Root causeProcess design
ActionMicrobiologics notified customers on about 02/18/2020 via letter titled "Urgent Medical Device Recall." Instructions included to inspect inventory for affected products. If found, instructions included the following: 1. REVIEW lab procedures to understand how thisinformation affects your usage; 2. USE OR DISCARD depending on your lab procedures and how this information affects your usage; 3. COMPLETE the response form provided; 4. RETURN the response form to recall@microbiologics.com; 5. KEEP this letter for your records; 6. CONTACT Microbiologics if a replacement kit is needed. Questio
Quantity3 units
DistributionWorldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K861022KWIK STIK(TM) LYFO(TM) DISK MICROORGANISMS
Micro-Bio-Logics, Apr 4, 1986