Recall Z-0587-2022 ยท JTR Kit, Quality Control For Culture Media
Microbiologics Inc
QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms
| Date | Feb 8, 2022 (initiated Feb 18, 2020) |
| Status | Terminated |
| Reason | Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release. |
| Root cause | Process design |
| Action | Microbiologics notified customers on about 02/18/2020 via letter titled "Urgent Medical Device Recall." Instructions included to inspect inventory for affected products. If found, instructions included the following: 1. REVIEW lab procedures to understand how thisinformation affects your usage; 2. USE OR DISCARD depending on your lab procedures and how this information affects your usage; 3. COMPLETE the response form provided; 4. RETURN the response form to recall@microbiologics.com; 5. KEEP this letter for your records; 6. CONTACT Microbiologics if a replacement kit is needed. Questio |
| Quantity | 10 units |
| Distribution | Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japa |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K861022 | KWIK STIK(TM) LYFO(TM) DISK MICROORGANISMS Micro-Bio-Logics, Apr 4, 1986 |