ClearanceBrief
Recall Z-0616-2022 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

GE Healthcare MR superconducting magnets, a component of GE Healthcare Contour System, nuclear magnetic resonance imaging system.

DateDec 23, 2021
StatusOpen, Classified
ReasonThe magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
Root causeProcess control
ActionThe recalling firm issued letters dated 12/23/2021 via FedEx explaining the safety issue and the actions to be taken by the customer/user. The actions say that the customer can continue to use the system as normal but they are to follow the instructions listed in the letter and to ensure the system is inspected either by GE Healthcare or the organization who installed their ventilation system to confirm it complies with the safety requirements for ventilation. The customer was requested to complete the workflow which was to be accessed through a QR code or link in the letter within the Medic
Quantity17,228 devices total in this field correction
DistributionDistribution was nationwide, including Puerto Rico and the Virgin Islands. There was also government and military distribution. Foreign distribution was made to Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaidjan, Bahrain, Bangladesh, Belarus, Belgium, Bhutan, Bolivia, Bosnia and
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K950973MAGNETIC RESONANCE DIAGNOSTIC SYSTEM
GE Medical Systems, May 11, 1995
K964370MAGNETIC RESONANCE DIAGNOSTIC SYSTEM
GE Medical Systems, Jan 10, 1997