Recall Z-0479-2022 ยท IYE Accelerator, Linear, Medical

PTW NORTH AMERICA CORPORATION

Software BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032

DateAug 27, 2021
StatusTerminated
ReasonWhen a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or too-high measured absolute dose which affects patient plan verification results
Root causeSoftware design
ActionPTW US notified customer Urgent Safety Notice via email and telephone on 8/27/20. Letter states reason for recall, health risk and action to take: PTW will provide free-of-charge software updates. Users who use BeamAdjust 2.2/ VeriSoft 8.0 according to the above described application must update the software to VeriSoft 8.0.1/BeamAdjust 2.2.1 prior to further use of the software. For all other users PTW recommends using BeamAdjust 2.2.1/VeriSoft 8.0.1 too. Until the software updates are available PTW recommends avoiding the air density correction method. As a workaround, if air density corr
Quantity1 unit
DistributionUS Nationwide distribution in the state of CO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160405OCTAVIUS I, OCTAVIUS II, OCTAVIUS III, OCTAVIUS 4D
Ptw-Freiburg Physikalisch-Technische-Werkstaetten, Oct 20, 2017