Recall Z-0500-2022 ยท LLZ System, Image Processing, Radiological

Konica Minolta Healthcare Americas, Inc.

ImagePilot Versions 1.92 and 1.93-software primarily facilitates processing and presentation of medical images on display monitors for medical diagnostics Product number: ADFR and ADJC

DateDec 3, 2021
StatusTerminated
ReasonlmagePilot versions 1.92 and 1.93 paired with an AeroDR or Momentum Panel potential risk of incorrect length measurement/area calculation if incorrect power sequence is followed
Root causeUnder Investigation by firm
ActionKonica Minolta issued Urgent: Medical Device Recall lmagePilot Versions 1.92 and 1.93 Letter to Dealers on 12/8/21 and Customers on 12/15/21 via Fed'X. Letter states reason for recall, health risk and action to take: To ensure the correct sampling pitch is obtained from the AeroDR or Momentum panel, please refer to the enclosed "lmagePilot & AeroDR Start-up Sequence and Notice before Shutting Down lmagePilot." Konica Minolta has issued a countermeasure to minimize the risk of a potential incident. This corrective action can only be carried out on site by a Konica Minolta trained representativ
Quantity30 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K210066ImagePilot