ClearanceBrief
Recall Z-0654-2022 ยท IZI System, X-Ray, Angiographic

Shimadzu Medical Systems

Digital Angiography System Bransist safire Generator Model D150GC-40

DateDec 16, 2021
StatusOpen, Classified
ReasonDue to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.
Root causeProcess change control
ActionOn 12/13/2021, the firm sent an "Urgent: Voluntary Medical Device Recall Notice" to distributors who hand-delivered to customers informing them that there is a potential issue with the adjustment procedure for the calibration of the x-ray generators for the Digital Angiography Systems and occurs during the calibration of the fluoroscopic mode which could cause x-ray radiation dose rate to exceed regulation and/or standards. Customers are instructed to: -Check their systems for the affected generators. -Contact their distributor to schedule a system update and calibration. -If customers feel t
Quantity18
DistributionU.S.: CA, FL, GA, IL, IN, KY, LA, MI, MS, OH, PA, SC, TN, TX and WA O.U.S.: Not provided.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062360BRANSIST SAFIRE
Shimadzu Corp., Oct 5, 2006