ClearanceBrief
Recall Z-0553-2022 ยท OOI Real Time Nucleic Acid Amplification System

Luminex Corporation

ARIES HSV 1&2 Assay, Catalog/REF# 50-10017

DateNov 17, 2021
StatusTerminated
ReasonImpacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.
Root causeNonconforming Material/Component
ActionOn 11/17/2021, the firm notified affected customers via letter titled, "URGENT: MEDICAL DEVICE RECALL ARIES HSV 1&2 Assay." The letter indicated, if you observe any leaks, please contact Luminex Global Support Services at support@luminexcorp.com. Please inform all those who need to be aware within your organization of this potential product performance issue. Please contact Luminex Global Support Services at support@luminexcorp.com at your earliest convenience if you have a Kit with an affected lot to arrange for its replacement, and to confirm the disposal of any remaining Kits in your inve
Quantity329
DistributionWorldwide distribution - US Nationwide distribution in the states of New Jersey, California, Connecticut, Minnesota, Missouri, New Jersey, New Mexico, Pennsylvania, Tennessee, Washington, and Wisconsin) and the country of Malaysia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161495ARIES MI System, ARIES HSV 1&2 Assay Cassettes-Carton of 24 (IVD), ARIES HSV 1&2 Assay Protocol File Kit (IVD)
Luminex Corporation, Jun 30, 2016