Recall Z-0501-2022 ยท BSZ Gas-Machine, Anesthesia

Draeger Medical, Inc.

Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300

DateDec 16, 2021
StatusTerminated
Reason26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.
Root causeProcess control
ActionOn December 17, 2021, the firm notified customers via Urgent Medical Device Correction Notice. Customers were instructed to do the following: Please ensure that all users of the above-mentioned products and other concerned persons within your organization are made aware of this Urgent Medical Device Correction Notice. If you have provided the products to third parties, please forward a copy of this information. Please keep this information at least until the corrective measure has been completed. Please also complete and return the attached Customer Acknowledgment Card to confirm to us
Quantity5
DistributionUS Nationwide distribution in the states of NY, PA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072884FABIUS MRI
Draeger Medical AG & Co. KG, Jul 30, 2008