ClearanceBrief
Recall Z-0483-2022 ยท PHX Shoulder Prosthesis, Reverse Configuration

DePuy Orthopaedics, Inc.

GLOBAL UNITE STD STEM SZ 6 Part Number: 11000610

DateDec 7, 2021
StatusTerminated
ReasonStems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage
Root causeEquipment maintenance
ActionDePuy notified US hospital / government consignees on 12/7/21 letter Urgent Medical Device Recall (Removal) via the sales consult and product will be removed/notice delivered by hand or if not able to enter facility the form will be emailed and returns will be arranged. US sales/distributor offices will be notified via email. Affiliates outside of the US affiliates will be notified via the electronic system and recall activities will be carried out per local regulations. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately to determine if you
Quantity28 units
DistributionWorldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Netherlands, New Zealand, Norway, Poland, South Africa, Sweden, Switzerland, Thailand, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101996DEPUY GLOBAL UNITE SHOULDER SYSTEM
Depuy(Ireland), Dec 15, 2010
K170748GLOBAL UNITE Platform Shoulder System
Depuy Ireland UC, Jan 11, 2018