Recall Z-0441-2022 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

ARTISTE with syngo RT Therapist-linear accelerator systems to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer. Model Number: 8162815

DateDec 9, 2021
StatusTerminated
ReasonPotential safety issue under specific preconditons that may result in a user selecting the wrong site for treatment with the possibility to deliver dose to the wrong isocenter which could result in serious patient injury
Root causeOther
ActionSiemens issued Urgent Medical Device Corrction Letter CSAN TH001/21/S on 12/09/2021 to inform them of the potential issue and associated risk. Acton to take: 1. Carefully verify all planned treatment parameters against those actually set during patient setup before treating a patient. 2. Choose one of the following workflow options: - Use absolute table positions. In this case, positioning on a different isocenter would not match the planned table values and the system would prompt an override request. This minimizes the risk of selecting a wrong site fraction group. Or - In the case of mulit
Quantity13 units
DistributionUS Nationwide distribution in the states of IA, LA, NE, NJ, NV, NY, OH, PA, SC, SD.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142434ARTISTE SOLUTION, SYS_VC10C, PHASE 2 UPDATE WITH THE SYNGO RT THERAPIST & RT ONCOLOGIST WORKSPACES, V4.3.1_MR1
Siemens Medical Solutions USA, Inc., Nov 21, 2014