Recall Z-0530-2022 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product code 35700BAX2.

DateJan 29, 2022 (initiated Dec 29, 2021)
StatusOpen, Classified
ReasonThere is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.
Root causeDevice Design
ActionThe firm sent URGENT SAFETY COMMUNICATION letters on 12/29/2021 via first class mail, addressed to Directors of Biomedical Engineering, Risk Management, Nursing, and Nurse Educators. The letter described the problem, the affected product (Spectrum V8 and Spectrum IQ infusion pumps), hazard involved, and the actions to be taken by customers. Acknowledge receipt on customer portal https://BaxterFieldActionCustomerPortal.onprocess.com/. Follow on-screen instructions, infusion setup instructions in the Preparing the Pump and IV Sets and Programming the Pump sections, and upstream occlusion alar
Quantity146,502 units
DistributionWorldwide Distribution. US nationwide Puerto Rico and St. Thomas. There was government/military distribution. Countries of Canada, Bahamas, Barbados, Bermuda, Guyana, Jamaica, and Trinidad and Tobago.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133801BAXTER SIGMA SPECTRUM INFUSION PUMP WITH MASTER DRUG LIBRARY
Baxter Healthcare Corporation, May 7, 2014