Recall Z-0492-2022 ยท OWB Interventional Fluoroscopic X-Ray System

Canon Medical System, USA, INC.

Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-8000F, and INFX-8000H.

DateDec 20, 2021
StatusOpen, Classified
ReasonAbnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel.
Root causeVendor change control
ActionOn 12/21/2021, the firm notified affected customers via a letter titled, "URGENT: MEDICAL DEVICE CORRECTION." The letter indicated, "Corrective Action: When the new parts become available, your Canon service representative will contact you for an appointment to schedule the installation of the modified x-ray flat panel detector and will install on your system. Requests to Customers: Should this problem occur before the modified x-ray flat panel detector is installed on your system, please attempt computer reboot, and contact your Canon service representative. If you have any questions regar
Quantity8
DistributionUS Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181670Alphenix, INFX-8000V/B, V8.0
Canon Medical Systems Corporation, Jul 25, 2018
K181804Alphenix, INFX-8000C/B, V8.0
Canon Medical Systems Corporation, Aug 2, 2018
K182415Alphenix, INFX-8000F/B, V8.0
Canon Medical Systems Corporation, Sep 28, 2018
K182546Alphenix, INFX-8000H/B, V8.0
Canon Medical Systems Corporation, Oct 17, 2018