ClearanceBrief
Recall Z-0519-2022 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic

GE Healthcare, LLC

GE Vscan Extend, being sold as "Vscan Extend Dual DICOM"

DateJan 24, 2022 (initiated Nov 4, 2021)
StatusTerminated
ReasonThe firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
Root causeDevice Design
ActionThe firm distributed an Urgent Medical Device Correction letter to customers on 11/17/2021 informing customers of the issue with Vscan Extend devices not powering on because of depleted internal battery. This issue occurs if the device is not turned off prior to the external battery being removed or when the device is connected to a personal computer without connecting to AC power. The actions needed to be taken by the user until a software update could be deployed were identified in this notice as: make sure to power off the Vscan Extend device before removing the external battery, always con
Quantity12,526 total reported units (9,771 units of this model)
DistributionDomestic: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT. NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, PR, and DC. International: Albania, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia, Braz
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161588Vscan Extend
GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC, Aug 31, 2016
K180995Vscan Extend
GE Medical Systems Ultrasound and Primary Care Diagnostics, Jun 25, 2018