ClearanceBrief
Recall Z-0475-2022 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SOMATOM Definition Edge -Computed tomography system Model 10590000

DateNov 19, 2021
StatusOpen, Classified
ReasonSoftware version VB20_SP5 my lead to a relevant degradation of head image quality, a potential risk to patients is misdiagnosis.
Root causeSoftware Design Change
ActionSiemens issued Urgent Medical Device Correction letter CSAN CT067/21/S on November 19, 2021, to notify affected customers of the head image degradation issue and to provide them with a workaround solution. Letter states reason for recall, health risk and action to take: To avoid potential misdiagnosis, it is mandatory to change to the new cupping correction setting CC = 4 immediately, restart the application, and (if applicable) reconstruct the raw data again. Cupping correction setting CC = 4 is a further development of CC = 1: The Image impression is very similar to setting CC = 1 but an imp
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190578SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM Definition AS Open, SOMATOM Edg
Siemens Medical Solutions USA, Inc., Jun 27, 2019