ClearanceBrief
Recall Z-0467-2022 ยท INM Chair, With Casters

Ki Mobility Llc

Ki Mobility, Little Wave Tyke (Weight Limit: 95 lbs), Specifically the IFU The Little Wave Tyke is a manually operated device that can be used for medical purposes and is used to assist a disabled person to move from one place to another while sitting. Intended to be used by persons up to 95 pounds for transport.

DateNov 19, 2021
StatusTerminated
ReasonThe Instructions for Use provided with the device contained information intended for use of a mechanical wheelchair, not a chair with casters.
Root causeProcess design
ActionThe firm distributed a Medical Device Correction Notice dated 12/07/2021 by first class mail to consignees. The notificaiton identified the reason for recall as the Owner's Manual/Instructions for Use for the Little Wave Tyke contained information intended for a mechanical wheelchair, not a mechanical chair with casters. Customers were instructed to distribute the provided updated Little Wave Tyke Owner's Manual/Instructions for Use (DCN0569.2) to end users, and if available request that customers destroy or dispose of the previous versions of the manual (DCN0569.0 and DCN0569.1). The updated
Quantity33 units shipped with affected manuals
DistributionDomestic: IL, IN, KY, MI, MO, NC, SC, and TN.