Recall Z-0527-2022 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Velano Vascular

PIVO Blood Collection Device 20G, REF: 202-0005

DateNov 10, 2021
StatusTerminated
ReasonIncomplete sealing of blood collection device unit packages may produce a breach in the sterile barrier enclosing the product. This could lead to microbial contamination of the device prior to use. The health effects range from no clinical effect up to bloodstream infection.
Root causeProcess control
ActionOn 11/10/21, recall notices were sent to Medical Directors, Risk Managers, Medical Device Safety Officers, Nurse Managers. 1) Locate and destroy all product subject to the recall following your institution s process for destruction. 2) If the affected product was previously used in patient care and no signs of infection are apparent, no additional steps need to be taken. If the affected product was used in patient care and signs of infection are present, follow your institution s established procedures. 3) Share this recall notification with all users within your facility network of the produc
Quantity123,700
DistributionUS Nationwide distribution in the states of CT, CO, UT, PA, IL, WA, CA, NJ, MI, FL, NC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193569PIVO(TM)
Velano Vascular, Jan 21, 2020