Recall Z-0458-2022 ยท EZY Device, Incontinence, Mechanical/Hydraulic

Uromedica Inc.

UROMEDICA Implantation Instrument Set (Sheath and Sharp Trocar), REF 750041, nonsterile.

DateNov 22, 2021
StatusTerminated
ReasonThe U-channel sheaths may cause damage to ProACT implant devices when the devices are passed through the sheath during a ProACT implantation procedure.
Root causeNonconforming Material/Component
ActionLetters dated 11/22/2021 were issued via email to the consignees explaining the reason for recall. The consignee is requested to identify and remove the affected u-channel sheath and sharp trocar from circulation at their facility. The consignee will be contacted by a Uromedica representative who will schedule an in-person visit and collect the affected U-channel sheath and sharp trocar. Once the instruments have been removed from the facility, replacement instruments will be provided.
Quantity18 devices
DistributionUS Nationwide distribution in the states of AZ, CA, CO, KS, MD, TX, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.