Recall Z-0495-2022 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Covidien Llc

LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837

DateJan 13, 2022 (initiated Dec 6, 2021)
StatusTerminated
ReasonCustomers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding, unintended tissue injury or loss, and delay in treatment while an alternate device is obtained.
Root causeDevice Design
ActionURGENT MEDICAL DEVICE RECALL notification letters dated 12/2021 were sent to customers. Customer Action Required: 1. Please immediately quarantine and discontinue use of the affected item code with the associated lot number listed above. 2. Please return affected product as indicated below. All products from the affected item code and associated lot number must be returned. 3. If you have distributed the product item code and lot number listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not
Quantity504 devices
DistributionWorldwide Distribution in the countries of Australia, Austria, Belgium, Czech Republic, France, Germany, Ireland, Kazakhstan, Netherlands, Spain, United Kingdom. No US distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162941Ligasure Blunt Tip, Sealer/Divider, Nano-coated
Covidien, Nov 14, 2016