Recall Z-0478-2022 ยท OBO Tomography, Optical Coherence

Haag-Streit USA Inc

OCT-Camera 211 01 A3; Running software iOCT-Control 3 v1.4

DateNov 3, 2021
StatusOpen, Classified
ReasonMalfunction of the automatic laser beam shut-off, the OCT unit might not recognize whether the laser beam is safely switched off
Root causeSoftware design (manufacturing process)
ActionHaag-Streit USA send out field safety notice 2021-10/01 on November 03, 2021. Actions to be taken by the customer/user: Should one or more continuous straight lines be displayed on the screen of the OCT camera during an operation on the posterior segment of the patient's eye (see picture in the notification), the described risk possibly exists. 1. Switch off the OCT camera immediately (menu OCT -> Switch off). 2. The OCT camera is not permitted to be used any more. Please contact your responsible country representative. 3. Perform the software update of the OCT camera. You will receive the n
Quantity3
DistributionUS Nationwide distribution in the states of OH, IN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142953OCT-Camera
Optomedical Technologies GmbH, Mar 4, 2015