Recall Z-0692-2022 ยท JGJ Photometric Method, Magnesium

Beckman Coulter Inc.

AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850

DateDec 8, 2021
StatusTerminated
ReasonLipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is only to the clinical interpretation of magnesium results in the presence of lipemia.
Root causeUnder Investigation by firm
ActionOn 12/08/2021, the firm sent recall notifications to customers via mail and e-mail. Customers with additional questions were encouraged to contact Customer Support: http://www.beckmancoulter.com or 1-800-223-0130 Action to be taken: "Retain a copy of this letter as it serves as current labeling. Where LIH influence check settings are enabled, customers must update the lipemia influence check settings on their AU/DxC AU analyzers. To determine if LIH influence check settings are enabled on your analyzer for Magnesium and to update your lipemia interference settings, perform the following acti
Quantity181,776 Kits
DistributionUS: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New