Recall Z-0528-2022 ยท PGN Growing Rod System- Magnetic Actuation

NuVasive Specialized Orthopedics, Inc.

MAGEC 1 System

DateJun 25, 2019
StatusOpen, Classified
ReasonSafety Alert to all surgeons informing them of post-implantation fractures of an internal metallic component for device manufactured prior to March 26, 2015.
Root causeDevice Design
ActionBeginning on 06/25/2019, a Field Safety Notice was communicated to all surgeon informing them that post-implantation fracture of an internal metallic component (i.e., locking pin) has been observed in early versions of the MAGEC System Rods that were manufactured prior to March 26, 2015. Fracture of the locking pin may affect the ability of the device to lengthen and may be associated with Titanium wear debris generation and localized tissue discoloration. Recommended User Actions: - NuVasive, Inc. voluntarily issues this Safety Alert to reinforce the Instructions for Use (IFU) of the MAGEC
Quantity3,056 devices
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140178ELLIPSE MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Ellipse Technologies, Inc., Feb 27, 2014