Recall Z-0471-2022 ยท OWB Interventional Fluoroscopic X-Ray System

Philips North America Llc

Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue; Azurion R2.2 emergency stop recovery issue Model Numbers: 722221, 722228, 722225, 722068, 722226, 722078, 722224

DateNov 12, 2021
StatusTerminated
Reason(1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not executed correctly and movement functionality is not available
Root causeSoftware design
ActionPhilips issued Field Safety Notice on 11/12/21 U.S.A, Customer notified via certified mailing from the United States Postal Service (USPS). Letters outside of the U.S.A. will be distributed through the Philips Markets Organization. Letter states reason for recall, health risk and action to take: Start-up problem: If the start-up problem occurs, perform a Cold Restart* of the system. b) Emergency stop recovery problem: If the emergency stop recovery problem occurs, perform a new Warm Restart* of the system. c) Place this Field Safety Notice with the documentation of the system until Philips h
Quantity10 units (1 US and 9 OUS)
DistributionWorldwide distribution - US Nationwide distribution in the state of WA and the countries of France, Germany, Japan, Poland, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181830Azurion R2.0
Philips Medical Systems Nederland B.V., Aug 2, 2018