Recall Z-0462-2022 ยท MZP Assay, Hybridization And/Or Nucleic Acid Amplification For Detection Of Hepatitis C Rna,Hepatitis C Virus

Abbott Molecular, Inc.

Alinity m System, Part No. 08N53-002

DateDec 6, 2021
StatusTerminated
ReasonThere is a software defect that can cause the Amplification Detection clamp CLEAN position to be set too high.
Root causeSoftware design
ActionOn December 06, 2021 Abbott Molecular issued a Product Recall letter to all customers who received an Alinity m instrument (List Number 08N53 002). Customers are instructed to follow the Necessary Actions outlined in the Customer Letter until an Abbott Molecular representative adjusts the amplification detection clamp height on the Alinity m System. Abbott personnel will follow up with customers by e mailing and or follow up visits/phone calls.
Quantity742 devices
DistributionWorldwide distribution - US distribution and the countries of Australia, Austria, Belgium, Brazil, Cambodia, Canada, Chile, Colombia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Malawi, Malaysia, Mexico, Montenegro, Morocco, Mozamb
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.