ClearanceBrief
Recall Z-0460-2022 ยท MZP Assay, Hybridization And/Or Nucleic Acid Amplification For Detection Of Hepatitis C Rna,Hepatitis C Virus

Abbott Molecular, Inc.

Alinity m System, Part No. 08N53-002

DateDec 6, 2021
StatusOpen, Classified
ReasonThere is a potential for abnormal (non-sigmoidal) amplification curves.
Root causeUnder Investigation by firm
ActionOn December 06, 2021 Abbott Molecular issued a Product Correction letter to all customers who received an Alinity m instrument (List Number 08N53 002). Customers are instructed to follow the Necessary Actions outlined in the Customer Letter until an Abbott Molecular representative adjusts the customers clamp bar. Abbott personnel will follow up with customers by e mailing and or follow up visits/phone calls.
Quantity742 devices
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Cambodia, Canada, Chile, Colombia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Malawi, Malaysia, Mexico, Montenegro, Morocco, Mozambiq
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.