ClearanceBrief
Recall Z-0355-2022 ยท OWB Interventional Fluoroscopic X-Ray System

Philips North America Llc

BV Endura with Software Release 2.3- A mobile, diagnostic X-ray image acquisition and viewing system. Model Number: 718075

DateNov 12, 2021
StatusOpen, Classified
ReasonInstructions for Use do not specify the maximum surface temperature of the X-ray tank (assembly housing the X-ray tube, beam filter & cooling oil) and the image intensifier/detector of the system as required an result in burns
Root causeDevice Design
ActionPhilips issued an Urgent Medical Device Correction letter in the USA on 11/12/21 via Certified mailing United States Postal Service. Recalling outside the US is done through the local Philips organizations in the respective countries Letter states reason for recall, health risk and action to take: Pay attention to the Heat Indication Level Icons shown in the Viewing Station to know the temperature of the X-Ray Tank. o Follow the instructions provided in Annex I of this letter for a safe use of the system. o Circulate this notice to all users of this device so that they are aware of the iss
Quantity22 units US
DistributionWorldwide distribution - US Nationwide and the countries of Afghanistan, Algeria, American Samoa, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cameroon, Chile, China, Colombia, Cook Islands, Costa Rica, Croatia, Cuba, Cyprus, Czech Repu
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010435PHILIPS BV PULSERA/ENDURA
Philips Medical Systems, Inc., Mar 12, 2001