ClearanceBrief
Recall Z-0468-2022 ยท BSZ Gas-Machine, Anesthesia

Getinge Usa Sales Inc

Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: Flow-c Anesthesia System 6887700 07325710009765 Flow-e Anesthesia System 6887900 07325710010457 Flow-i Anesthesia System C20 6888520 07325710010617 Flow-i Anesthesia System C30 6888530 07325710010624 Flow-i Anesthesia System C40 6888540 07325710010631 Flow-i C20 Anesthesia System 6

DateDec 9, 2021
StatusOpen, Classified
ReasonA potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride.
Root causeUnder Investigation by firm
ActionOn December 9, 2021, the firm began distribution of Urgent Medical Device Recall letters to affected customers. Customers were informed of the product issue and potential risks to health, which may include irritation of respiratory tract and in worst case may lead to lung edema and/or severe hypocalcemia which may be delayed for 24-48 hours after exposure. Actions to be taken by the Customer: " Please examine your inventory immediately to determine if you have any of the Vaporizer Sevoflurane, Maquet Filling with the product codes/serial numbers listed in this notice. " Should you have any
Quantity50 US; 898 OUS
DistributionWorldwide distribution - US Nationwide distribution in the states of VA and MO. The countries of Austria, Bahrain, Bolivia, Brazil, Bulgaria, Chile, China, Colombia, Estonia, Iraq, Israel, Italy, Japan, Kenya, Malaysia, Mexico, Morocco, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Poland, Port
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia System
Maquet Critical Care AB, Mar 18, 2020