ClearanceBrief
Recall Z-0447-2022 ยท ITI Wheelchair, Powered

Invacare Corporation

Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: SRX-20R, SRX-20SP, SRX-20MP

DateNov 12, 2021
StatusOpen, Classified
ReasonPower Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury
Root causeSoftware design
ActionInvacare issued Urgent Medical Device Letter issued on 11/12/21 to consignees who received devices with LiNX Gyro modules containing firmware version 6.1.2. This letter also contains instructions for consignees who upgraded devices in the field to version 6.1.2. A Consumer Letter will be provided on the recall website for Providers to send to end users who are impacted. Letter states reason for recall, health risk and action to take: Visit www. Invacare.com/RC916 within five (5) business days of receiving this notice. Enter the requested identification information to access the list of power
Quantity1 unit US
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K202379Invacare AVIVA Storm RX Power Chair
Invacare Corporation, Mar 19, 2021