Recall Z-0476-2022 ยท OWB Interventional Fluoroscopic X-Ray System

Philips North America Llc

Wireless Footswtich used with the following Philips Allura Xper and Azurion Interventional Fluoroscopic X-ray System . Model Number: 722001, 722002, 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722028, 722029, 722033, 722034, 722035, 722036, 722038, 722039, 722058, 722059, 722065, 722066, 722063, 722

DateJan 10, 2022 (initiated Oct 29, 2021)
StatusOpen, Classified
ReasonWireless foot switch connection issues causing interruption of Fluoroscopy and exposure
Root causeDevice Design
ActionPhilips notified customers via an Urgent Medical Device Correction letter on October 29th, 2021 via Certified Mail. Letters outside of the U.S.A. will be distributed through the Philips Markets Organizations. Letter states reason for recall, health risk and actions that should be taken by the user in order to prevent risks for patients or users " Ensure that the wired foot switch is always connected to the system. The wired foot switch is always provided with the Allura Xper and the Azurion systems. " Immediately start using the wired foot switch in case the wireless foot switch loses connec
Quantity1971 Total; USA: 681 systems R.O.W.: 1,290 systems
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Australia, Austria, Bangladesh, Belgium, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Ital
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161563Allura Xper FD series and Allura Xper OR Table series
Philips Medical Systems Nederland B.V., Jul 29, 2016
K200917Azurion R2.1
Philips Medical Systems Nederland B.V., May 1, 2020