Recall Z-0363-2022 ยท HRS Plate, Fixation, Bone

Smith & Nephew, Inc.

smith&Nephew EVOS MEDIAL DISTAL FEMUR PLATE, L, 5H, 90MM, REF 72573101

DateNov 1, 2021
StatusTerminated
ReasonTesting vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.
Root causeUnder Investigation by firm
ActionThe firm initiated the correction by email and letter on 1 NOV 2021. The notice explained the problem and requested that the consignees check their inventory and complete the Inventory Correction Certificate form. Distributors were directed to notify their customers of the field action and ensure that these actions are carried out.
Quantity38 units
DistributionUS Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201918EVOS Large Fragment Plating System
Smith & Nephew, Inc., Nov 16, 2020