Recall Z-0239-2022 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Philips Healthcare

DigitalDiagnost C90

DateSep 22, 2021
StatusOpen, Classified
ReasonThe Aleva monitor requires certain labels to be applied, some of which are required by FDA Electronic Product Radiation Control performance standards. This includes the Radiation Warning Label [21CFR 1020.30 (j)] and the X-ray control certification statement [21 CFR 1010.2]. Philips has determined that some systems at customer locations did not have these labels applied.
Root causeRadiation Control for Health and Safety Act
ActionPhilips sent an Electronic Product Radiation Correction letter to inform customers. The letter identified the affected product, problem and actions to be taken. Customers were informed Philips Field Service Engineer will visit their site to check for the required labels and if they are missing, will resolve the issue by applying the required labels. Customers were instructed to complete and return the attached acknowledgment form to Philips DXR via email to DIFCO@philips.com. For questions contact Customer Care Solutions Center 1-800-722-9377.
Quantity281 in total
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K202564DigitalDiagnost C90
Philips Medical Systems Dmc GmbH, Sep 30, 2020