Recall Z-0319-2022 ยท HSB Rod, Fixation, Intramedullary And Accessories

NuVasive Specialized Orthopedics, Inc.

NUVASIVE PRECICE Intramedullary Limb Lengthening (IMLL) and (Short) System: Model Number/Item Number/Product Description: CPA1-000 PA0382-001 End Cap CPA2-000 PA1051-000 PACKAGED END CAP, 8.5/ 10.7MM, +0MM CPA2-005 PA1051-005 PACKAGED END CAP, 8.5/ 10.7MM, +5MM CPA2-010 PA1051-010 PACKAGED END CAP, 8.5/ 10.7MM, +10MM CPA2-015 PA1051-015 PACKAGED END CAP, 8.5/ 10.7MM, +15MM CPA2-020 PA1051-020 PAC

DateDec 1, 2021 (initiated Oct 29, 2021)
StatusOpen, Classified
ReasonThe Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.
Root causeDevice Design
ActionOn 11/01/2021, the firm emailed an "URGENT FIELD SAFETY NOTICE" to customers as a follow-up from it February and April 2021 communications. This notice informs of updates to the U.S. IFU which includes: a. Clarifying that - The PRECICE Intramedullary Limb Lengthening System is intended for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones in adults. b. Warnings - Patients of the PRECICE Intramedullary Limb Lengthening System should not be implanted with more than two devices at a time, and the patient s weight should
QuantityTotal 11,861 devices (IMLL=10,657 and Short=1,204)
DistributionU.S.: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, TN, TX, UT, VA, WA, WI, and WV. O.U.S.: Argentina, Australia, Austria, Belgium, Canada, Colombia, Chile, Cyprus, Denmark, F
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101997ELLIPSE INTRAMEDULLARY LIMB LENGTHENING SYSTEM
Ellipse Technologies, Inc., Jul 12, 2011
K113695ELLIPSE PRECICE TRAUMA NAIL SYSTEM
Ellipse Technologies, Inc., Mar 28, 2012
K141739RESIDUAL LIMB LENGTHENING SYSTEM
Ellipse Technologies, Inc., Oct 7, 2014
K151131PRECICE Intramedullary Limb Lengthening System
Ellipse Technologies, Incorporated, Jun 19, 2015