ClearanceBrief
Recall Z-0439-2022 ยท QKO Reagent, Coronavirus Serological

bioMerieux, Inc.

VIDAS SARS-COV-2 IgG, Ref 423834-01

DateSep 22, 2021
StatusOpen, Classified
ReasonbioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to potential delayed results as the user must run another test.
Root causeNonconforming Material/Component
ActionFor products with No remaining shelf life (taking into account the revised expiration date): - Identify impacted lots - Stop using the listed impacted lots - Destroy the remaining impacted lots in your inventory For products with remaining shelf life (taking into account the revised expiration date): - Identify impacted products and lots listed - Update product expiration date per your internal procedures of the remaining usable lots to meet the new revised expiration date specified in Table 1 - Continue product use until the revised expiration date. Consignees are asked to distribute this
QuantityTwo lots
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.