ClearanceBrief
Recall Z-0346-2022 ยท CBK Ventilator, Continuous, Facility Use

Covidien Llc

Puritan Bennett 980 Series Ventilator, 980X1ENDIPC

DateDec 9, 2021 (initiated Nov 4, 2021)
StatusTerminated
ReasonManufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Root causeReprocessing Controls
ActionOn November 4, 2021, the firm began to notify customers through Urgent Medical Device Correction letters. Customers were informed of the device defect that may lead to the ventilator becoming inoperable during use. Actions you should take Immediately discontinue use of the affected ventilators. Remove the affected ventilators from clinical service and quarantine the affected ventilators until such time that a Medtronic Technical Service Engineer inspects and replaces the affected printed circuit board assemblies. Notify all personnel in all care environments in which the affected PB980 se
Quantity278 total
DistributionWorldwide distribution - US Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193056Puritan Bennett 980 Series Ventilator System
Covidien, LLC, Nov 20, 2020