Recall Z-0379-2022 ยท FDT Duodenoscope And Accessories, Flexible/Rigid

Olympus Corporation of the Americas

EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V

DateDec 10, 2021 (initiated Nov 2, 2021)
StatusOpen, Classified
ReasonUse of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to patients.
Root causeUnder Investigation by firm
ActionOlympus issued Urgent Medical Device Corrective Action on 11/02/21. Letter states reason for correction, health risk and action to take: 1. Consider transitioning from TJF-Q180V to a next generation duodenoscope through the Olympus incentive program. 2. Increase preventative maintenance frequency from annual to 6 months. 3. Evaluate existing duodenoscopes: https://medical.olympusamerica.com/TJFQ180VChecklist 4. The Technical Assistance Center (TAC) can assist you in answering questions on the TJF-Q180V inspection steps. 1-800-848-9024, option 1. 5. Use in-service training, the Olym
Quantity14,447 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143153EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V
Olympus Medical Systems Corp., Jan 15, 2016