Recall Z-0291-2022 ยท MIH System, Endovascular Graft, Aortic Aneurysm Treatment
Medtronic Vascular, Inc.
Endurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
| Date | Oct 15, 2021 |
| Status | Open, Classified |
| Reason | Due to, during stent graft deployment, the spindle may detach from the distal end of the spindle hypotube. |
| Root cause | Under Investigation by firm |
| Action | On or about 10/15/2021, the firm initiated a global verbal notification to customers informing them that is the potential for a specific subset of 18Fr Stent Graft Systems to have the potential for the spindle to detach from the spindle hypotube during deployment. On 10/18/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS 2-day delivery. Customers are instructed to take the following actions: 1) Return all unused affected devices to the Recalling Firm via their sales representative or calling customer service at 1-888-283-7868 or email rs.aorticorders@medtroni |
| Quantity | 511 devices |
| Distribution | Worldwide distribution - U.S. Nationwide distribution in the states of CA, FL, GA, MN, MO, NC, NJ, NM, NV, NY, PA, SC, TN, TX, VA, and WV. The countries of Austria, Belgium, Bulgaria, China, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hungary, Ireland, Italy, Japan, Jordan, Lithuania, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.