ClearanceBrief
Recall Z-0301-2022 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis Q ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848281

DateOct 20, 2021
StatusTerminated
ReasonIf SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Root causeSoftware design
ActionSiemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice dated 10/202/1 to customers via AX070/21/S, Siemens will update the system software via Update Instruction AX069/21/S. Letter states reason for recall, health risk and action to take: To avoid the problem, it is recommended to position the SID first and then release x-ray. Siemens will correct the software error (software VD12) via Update Instruction AX069/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organizatio
Quantity67 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181407Artis zee/zeego & Artis Q/Q.zen
Siemens Medical Solution USA, Inc., Aug 15, 2018